Re: News on the preAmp + CDC
Hall D Electronics:
Hi Yves,
It looks like you and Gerard are making quite a lot of progress with the
FDC. The pulses look reasonable but there is still some work left in
trying to understand the source of the oscillations you observed before.
I also looked at the drift time spectra at different sampling rates that
you obtained. The time structure of the main pulse is similar and of the
order of 200 ns. There shouldn't be any aliasing as the pulse structure
out of the preamp is already over 10 ns. However, there seems to be some
reflection or after pulsing, as the 200 MSPS picture clearly shows. If
you are using the shaper, check if this is not an artifact from the
shaper. Gerard may have chosen the shaping constants based on the CLAS
preamp. I and Simon observed a similar characteristic with the anode
shaper on the FDC.
Regards,
Fernando